1 overview, 1 structure and function, 2 recommended fields of application – Xylem TetraCon 925 User Manual
Page 4: Overview, Structure and function, Recommended fields of application, 1overview

Overview
TetraCon
®
925
12
ba75824e01
08/2009
1
Overview
1.1
Structure and function
Structure
Automatic sensor
recognition
The sensor electronics with the stored sensor data is in the connecting head.
The data include, among other things, the sensor type and series number.
With each calibration, the calibration data is written in the sensor and the
calibration history is recorded. The data is recalled by the meter when the
sensor is connected and is used for measurement and for measured value
documentation. Storing the calibration data in the sensor ensures that the
correct cell constant is automatically used if the sensor is operated with
several meters.
The digital transmission technique guarantees the failure-free
communication with the meter even with long connection cables. If the sensor
firmware is enhanced by WTW, it can be updated via the meter.
1.2
Recommended fields of application
z
On site measurements in rivers, lakes and wastewater
z
Fish farming
z
Ground water measurements
z
Applications in water laboratories
1
Voltage electrode (inside, 2x)
2
Current electrode (ring, 2x)
3
Temperature sensor in graphite enclosure
4
Shaft
5
Connecting head with active electronics
4
1, 2 3
5