The following steps are carried out at the photometer. A PC connection is no
longer required.
The update process takes approx. three minutes. Then the photometer
restarts and carries out the self-test.
Note
If the photometer does not correctly start after a software update (e.g. no self-
test due to a power failure during the software update):
Press the
1
Call up the main menu with the
2
Using the F1 function key
[Settings], open the Settings
menu.
3
Using
▲><▼>
,
select the menu
itemSoftware/methods update
and press
4
Using
▲><▼>, select PC as the
source and press
The update process starts.
General setup
04/16/07 9:52
Language
Date/Time
Display settings
User managementg
Measured value memory
Software/methods update
Reset
Data transfer/Printer
Save all data on USB memory
Software/methods update
04/16/07 9:52
Software/methods update
Select source of update data:
USB memory
PC
Cancel