Ii. sensor specifications – Xylem WQ Conductivity Sensor User Manual
Page 4

Global Water
800-876-1172
• globalw.com
- 4 -
II.
Sensor Specifications
a. Specifications.
Output Type:
Two 4-20mA outputs
Accuracy:
Conductivity: +/-0.5% of Reading
Temperature: +/-0.2ºC (0.4ºF)
Cond. Range/Resolution 0-200uS
0.1uS
200-2000uS
1uS
2-20mS
0.01mS
20-200mS
0.1mS
200-2000mS
1mS
Temp. Range/Resolution -5ºC to +70ºC
0.01ºC
+23ºF to +158ºF
0.02ºF
Pressure:
35 psi maximum (25m/82ft H
2
O)
Immersion Depth
36mm (1.4 inches)
Voltage Requirements: 10-36 VDC
Supply Current:
20ma plus sum of both 4-20mA outputs
Warm Up Time:
15 seconds recommended
Temperature Response: 99% in <20 Seconds
Operating Temperature: -5°C to +70°C (23º to 158ºF)
Storage Temperature:
-20°C to +100°C (-4º to 212ºF)
Dimensions:
0.86” (22mm) diameter x 8” (202mm) long
Weight:
8oz plus cable
b. Connections:
Red:
10-36 VDC
Black:
Ground
White:
4-20 mA Conductivity Output
Green:
4-20 mA Temperature Output
Warning: Always turn off the power when connecting or disconnecting
the sensor or it could be damaged.