beautypg.com

Xi. maintenance, Xii. troubleshooting – Xylem WS705 IMPROVED COMPOSITE/DISCRETE WATER SAMPLER User Manual

Page 12

background image

Global Water

800-876-1172

globalw.com

- 12 -

XI. Maintenance

Sampler

The Global Water WS705 Water Sampler requires minimal maintenance. The

sampler enclosure is rainproof and rugged. Avoid exposure to extremely rough

usage. Routinely wipe the carrying case and control panel face, rinse the pickup

hose and debris strainer, and wash the sample bottle with mild soap and warm

water. Additional plastic and glass sample bottles, neoprene tubing for the

sampler pump, bottle caps/float switches, removable debris strainers, and

pickup hose can be purchased from Global Water (see Appendix D section).

Battery

If the pump runs slowly, this is an indication that the battery requires recharging.

Global Water recommends fully recharging batteries approximately once per

month. In addition, the battery should be recharged before any extended use.

NOTE: The battery life will be maximized if recharged before it drains below 11

volts.

To recharge the battery, plug the included charger

into the charging jack on the outside on the sampler

enclosure. The WS705 has a 3-Stage charge

controller installed inside the battery enclosure that

will prevent over-charging. To prevent damage to the

battery and controller, use only the supplied charger.

Additional batteries and battery chargers are available

from Global Water (see Appendix D).

XII. Troubleshooting

a. Check that the battery is firmly connected to the control panel and confirm

that it is fully charged.

b. Press the Pump Test button on the front panel. The test button should

always run the pump, regardless of the state of any of the inputs.

c. Turn the power switch off for 1-2 seconds then back on to reset the sampler.

d. Check the Status LED on the front panel.

i. The LED is off: If the water sensors are being used, check the Rain

and Water Sensor inputs. Sensors must be in contact with moisture

and unused inputs must have one of the shorting plugs supplied with

the sampler plugged into the jack. If the Trigger Input is being used,