Holding the Trigger Input active past the end of the sample will cause the pump to continue

until the Trigger Input is released. When the sample is complete, the pump reverses, then

stops.

Flow Proportional Sampling

a. To set the WS705 up to take flow proportional samples, a flow monitor must

be attached to the control panel. In this configuration, the sampler is

designed to take a sample every time a specific amount of flow has passed

through the flow monitor.

b. The flow monitor must provide a momentary switch closure (250ms or longer)

or a pulse of 4-30VDC, whenever a sample is to be taken. The output of the

flow monitor must be wired to a 2-wire RCA jack as shown in Appendix C.

The phone plug is then inserted into the Trigger In jack on the control panel.

c. Set the Sample Interval knob to External Trigger mode and the Sample Size

knob to the desired sample size. The sample will be taken, flow

proportionally, as directed by the flow meter.