beautypg.com

Xi. maintenance – Xylem WS700R REFRIGERATED WASTEWATER SAMPLER User Manual

Page 12

background image

Global Water

800-876-1172

globalw.com

- 12 -

XI. Maintenance

Sampler

a.

The Global Water WS700R Water Sampler requires minimal maintenance.

Avoid exposure to extremely rough usage. Routinely wipe the control panel

face and housing, rinse the pickup hose and debris strainer, and wash the

sample bottle with mild soap and warm water. Additional plastic and glass

sample bottles, neoprene tubing for the sampler pump, bottle caps/float

switches, removable debris strainers, and pickup hose can be purchased

from Global Water.

Refrigerator
f. Clean the door seal using mild soapy water. Never use harsh cleansers or

scouring pads to clean any part. Wash the inside with warm water and

baking soda solution (about 2 tablespoons of baking soda to a quart of water).

Rinse and dry.

g. Defrost whenever frost on the wall of the freezer compartment becomes ¼”

thick. Never use a sharp or metallic instrument to remove frost as it may

damage the cooling coils. A punctured cooling coil will cause serious

problems and void the warranty. Global Water recommends using a plastic

scraper.

Battery
h. If the pump runs slowly, this is an indication that the internal battery requires

recharging. We recommend fully recharging batteries approximately once per

month. In addition, the battery should be recharged before any extended use.

NOTE: The battery life will last longer if recharged before it drains below 10.5

volts.

b. To recharge the battery, plug the AC power adapter into the control panel and

then plug the charger into a wall socket, ensure that the pump does not run

during the charging period. Full recharge will take about 12 hours. Additional

batteries and AC power adapters are available from Global Water.