X. specifications – Xylem WS700R REFRIGERATED WASTEWATER SAMPLER User Manual
Page 11
Global Water
800-876-1172
• globalw.com
- 11 -
X. Specifications
Operating Temperature:
0° to +70°C
Size of unit:
Control Panel: 14 ¼”L X 4”W X 6 ¼”H
Refrigerator: approx 20” x 20” x 20”
Weight:
57lb (Shipping Weight 65lb)
Materials:
Enclosure:
Polycarbonate
Bottle:
2 gallon Polyethylene
Pickup Hose:
15' nylon reinforced 1/4" ID polyethylene
flexible tubing section with intake strainer
Pump Tubing:
Neoprene 1/4" ID, 7/16” OD
Sample Pump:
Flow Rate:
1000 ml per minute at a 4 foot head
Type:
Peristaltic
Maximum Lift:
~20 feet
Battery:
Rechargeable 5 AH Gel Cell
Battery Life:
~1 hour, continuous pumping under load
Standby: 3 months while still retaining
enough power to run the pump to capacity
Start Delay:
16 time settings from 0 to 12 hours
Composite Interval:
15 time settings from 5 min. to 12 hours
plus an External Trigger mode setting
Sample Size:
15 composite sample sizes from 50ml to
2 liters plus a Full Bottle discrete setting
(Approximate sizes at 4 foot head)
External Trigger Input:
250mS minimum pulse width
Switch closure or 4-24VDC
Pulse Output:
5VDC one-second pulse
1000ohm output impedance
Bottle Switch Input:
Switch closure Input
Floating read switch in bottle
Rain and Water Sensors:
Optional moisture sensors or switch
closure inputs
Internal Fuse:
10A Slow-Blow