beautypg.com

Xylem WS700 COMPOSITE/DISCRETE WATER SAMPLER User Manual

Page 11

background image

Global Water

800-876-1172

globalw.com



- 11 -

c.

Set the Sample Interval knob to External Trigger mode and the Sample Size

knob to the desired sample size. The sample will be taken, flow

proportionally, as directed by the flow meter.


X. Specifications

Operating Temperature:

0° to +70°C

Size of unit:

22”H X 17"W X 9"D

Weight:

20lb (Shipping Weight 22lb)

Materials:

Enclosure:

Expanded UV protected PVC

Bottle:

2 gallon Polyethylene

Pickup Hose:

15' nylon reinforced 1/4" ID polyethylene

flexible tubing section with intake strainer

Pump Tubing:

Neoprene 1/4" ID, 7/16” OD

Sample Pump:

Flow Rate:

1000 ml per minute at a 4 foot head

Type:

Peristaltic

Maximum Lift:

~20 feet

Battery:

Rechargeable 5 AH Gel Cell

Battery Life:

~1 hour, continuous pumping under load

Standby: 3 months while still retaining

enough power to run the pump to capacity

Start Delay:

16 time settings from 0 to 12 hours

Composite Interval:

15 time settings from 5 min. to 12 hours

plus an External Trigger mode setting

Sample Size:

15 composite sample sizes from 50ml to

2 liters plus a Full Bottle discrete setting

(Approximate sizes at 4 foot head)

External Trigger Input:

250mS minimum pulse width

Switch closure or 4-30VDC

Pulse Output:

5VDC one-second pulse

1000ohm output impedance