Maintenance

a. Global Water recommends verifying the sensor’s calibration with a

sounder or other measuring device once every 6 months.

b. The screen on the end of the sensor must be periodically checked for

clogging from mud, sludge, and other debris. Wash the screen with
clean water and/or scrub it gently with a toothbrush. Do not insert
objects through the screen, as this may cause damage to the sensor.

VI.

Trouble Shooting

Issue: Sensor reading incorrectly
a. Verify power source is supplying the correct voltage.

b. Verify that the vent tube has not been kinked or sealed. The sensor

uses this tube to compensate for barometric pressure changes.

c. Clean the sensor following the maintenance instructions.
d. Verify the sensor’s calibration.

Issue: Water in the vent tube
a. If water gets into the vent tube of the cable place it next to a heater for

24 hours to dry the inside of the cable.


Other issues
a. Call Global Water for tech support: 800-876-1172 or 916-638-3429

(many problems can be solved over the phone). Fax: 916-638-3270 or

Email:

[email protected]

.

When calling for tech support, please have the following information
ready;

1. Model #.
2. Unit serial number.
3. P.O.# the equipment was purchased on.

4. Global Water’s sales number or the invoice number.