Storage, Aging – YSI TruLine ORP 51 User Manual
Page 8

8
T 814246
04/2014
Storage, Aging
TruLine
Storage
During short
measuring breaks
Immerse the electrode in the reference electrolyte.
Combination
electrode
Reference electrolyte
YSI Model
(see page 11)
TruLine ORP 51
3 mol/l KCl, Ag
+
-free
KCl-250 (250 ml)
Prior to the next measurement, briefly rinse the combination electrode with
the test sample or deionized water.
Overnight or
longer
Insert the clean electrode into the watering cap filled with reference
electrolyte.
Note
During longer storage periods, salt deposits may develop on the watering
cap. They do not affect the measuring characteristics and can easily be re-
moved with deionized water when the combination electrode is put into oper-
ation again.
Aging
Every ORP combination electrode undergoes a natural aging process. Ex-
treme operating conditions can considerably shorten the lif etime of the com-
bination electrode. These are:
Strong acids or lyes, hydrof luoric acid, organic solvents, oils, fats, bro-
mides, sulf ides, iodides, proteins
High temperatures
Great changes in pH and temperature.
The warranty does not cover failure caused by measuring conditions and
mechanical damage.