Specifications – Xylem 407119A HOT WIRE ANEMOMETERS User Manual
Page 3

Model 407119A Version 4.1 October 2005
3
Specifications
General Specifications
Display
Dual function 5-digit LCD
Measurement units
Air Velocity: m/s, km/h, ft/min, knots, mph;
Air Flow: CMM (m
3
/min) and CFM (ft
3
/min);Temperature: °C and °F
Data hold
Freezes displayed reading
Sampling rate
Display update rate:1 second (approx.)
Sensors
Air velocity and temperature sensors: Thermistor type
MAX/MIN Memory
Record and view Maximum and Minimum readings
Average feature
Averages up to 20 readings
Automatic Power off Auto shut off after 15 minutes
Data Output
RS-232 PC serial interface with 16-bit data stream output
Operating Temp.
32 to 122°F (0 to 50°C)
Operating Humidity
Max. 80% Relative Humidity
Power Supply
Four (4) ‘AA’ 1.5V batteries or optional AC adaptor
Power Current
70mA DC (approx.)
Weight (meter only)
1.15 lbs. (521g) with batteries installed
Dimensions
Main instrument: 7.9 x 3.0 x 1.5" (200.0 x 76.2 x 36.8mm)
Telescoping Sensor: 0.5" (12.7mm) diameter head
Min length of sensor: 8” (260mm); Max. length: 37” (0.94m) with 5.5’
(1.7m) cable
Range Specifications
Air Velocity
Range
Resolution
Accuracy (of rdg)
m/s (meters per second)
0.2 – 17.0 m/s
0.1
±(5% + 5 digits)
km/h (kilometers per hour)
0.7 – 61.2 km/h
0.1
±(5% + 20 digits)
ft/min (feet per minute)
40 – 3346 ft/min
1
±(5% + 100 digits)
mph (miles per hour)
0.5 – 38.0 mph
0.1
±(5% + 10 digits)
knots (nautical miles per hour)
0.4 to 33.0 knots
0.1
±(5% + 10 digits)
Air Flow Measurements
Range
Resolution
Area Range
CMM (cubic meters per min.)
0 – 36,000 m
3
/min
0.001 to 1
0.001 to 30.0 m
2
CFM (cubic feet per minute)
0 – 1,271,200 ft
3
/min
0.01 to 100
0.01 to 322.91ft
2
Range Resolution
Accuracy
Air Temperature
32 to 122
o
F
(0 to 50
o
C)
0.1
o
F/C ±1.5
o
F (0.8
o
C)