Xylem FDO 925 User Manual
Page 15

FDO
®
925
Technical data
31
ba75898s01
04/2010
Measurement
conditions
Measuring ranges at 20 °C
(68 °F)
0 ... 20 mg/l D.O.
0 ... 200 % D.O
.
saturation
0 ... 400 mbar D.O. partial pressure
Temperature range
0 ... 50 °C (32 ... 122 °F)
Max. admissible overpres-
sure
2.5 x 10
5
Pa (2.5 bar)
Immersion depth
min. 6 cm
max. 25 m (depending on the cable length)
Operating position
any
Approach flow
not required
Storage conditions
Recommended storing
method
in the check and storage vessel,
FDO
®
Check
Storage temperature
0 ... 50 °C (32 ... 122 °F)
Characteristic data
on delivery
Zero signal
< 0.1 % of the saturation value
Response time at 20 °C
(68 °F) in stirred solution
t
90
(90% of the final value display after) < 30s
t
95
(95% of the final value display after) < 45s
t
99
(99% of the final value display after) < 60s
Response time of tempera-
ture measurement
t
99
(99% of the final value display after) < 60s
Precision of temperature
measurement
± 0.2 K
Working life of the sensor
cap
Min. 1 year with authorized use
- Lab 850 BENCHTOP pH METER (73 pages)
- Lab 850, up to serial number 09470314 (61 pages)
- Lab 860 BENCHTOP pH METER (74 pages)
- Lab 870 BENCHTOP pH METER (71 pages)
- Lab 960 (77 pages)
- Lab 970 (71 pages)
- CG 839 (8 pages)
- CG 842 (44 pages)
- CG 843P (72 pages)
- CG 843 (68 pages)
- CG 853P (76 pages)
- CG 853 (74 pages)
- CG 855 (11 pages)
- HandyLab 100 QuickStart (50 pages)
- HandyLab 100 (66 pages)
- HandyLab 200 QuickStart (50 pages)
- HandyLab 200 (52 pages)
- HandyLab 600 QuickStart (50 pages)
- HandyLab 600 (72 pages)
- HandyLab 680 QuickStart (50 pages)
- HandyLab 680 (120 pages)
- HandyLab 700(EX) (128 pages)
- HandyLab 700 (76 pages)
- HandyLab 780 (164 pages)
- handylab LF 11 (44 pages)
- handylab LF 12 (64 pages)
- handylab OX 12 (60 pages)
- handylab multi 12 (82 pages)
- handylab pH 11 (48 pages)
- handylab pH 12 (68 pages)
- handylab pH LF 12 (80 pages)
- IQ SensorNet FDO 70x IQ (SW) (48 pages)
- IQ SensorNet ViSolid 700 IQ (SW) (58 pages)
- MultiLab 4010-2 (112 pages)
- MultiLab 4010-3 (112 pages)
- EBI 40 TK-12 (18 pages)
- TFX 410 (38 pages)
- EBI 20 Serie (20 pages)
- EBI20 TEMPERATURE/HUMIDITY DATALOGGERS (40 pages)
- EBI 20-TF (40 pages)
- TFX 422 (36 pages)
- ST 1000 (42 pages)
- TLC 1598 (2 pages)
- TTX 100/110/120 (28 pages)
- EBI 25 (18 pages)