Xylem CL200 Waterproof Total Residual Chlorine Tester User Manual

Page 5

background image

CL200 V6.0 06/05

5

Storing Readings

1. Press the MODE button to store a reading. The storage location

number will be displayed followed by the stored reading and the
“HOLD” indicator.

2. Only one reading can be stored during the 120 second measurement

cycle.

3. Press the MODE button again to return to normal operation.

4. If more than 15 readings are stored, previously stored readings

(starting with number 1) will be overwritten.

Recalling Stored Readings

1. Press the CAL button and then press the MODE button immediately

after CAL is displayed; a location number (1 through 15) will flash.

2. The last stored reading taken will be displayed first. To scroll through

the stored readings, press the MODE button. The location number is
displayed first, followed by the reading stored in that location.

3. To exit the storage mode, press the CAL button and the CL200 will

return to normal operation.

Note: If the batteries are removed all stored readings will be lost.

Changing the Temperature Units

Press and hold the CAL button for approx. 3 seconds to change the
temperature units.

Auto-Power OFF

The auto power off feature automatically shuts the CL200 off 10 minutes
after the last keypress.

Low Battery Indication

When the batteries become weak the “BAT” icon will appear in the
display. Refer to the Maintenance section for battery replacement
information.