H-4191

SDI-12 Command and Response Protocol 2-5

2.4 Continuous Measurements

This is a new command for the Version 1.2 SDI-12 Specification. Sensors that are able to

continuously monitor the phenomena to be measured, such as a cable position, do not require a

start measurement command. They can be read directly with the R commands (R0!...R9!). The R

commands work exactly like the D (D0!...D9!) commands. The only difference is that the R

commands do not need to be preceded with an M command.

The H-4191

does not support the aR0! continuous measurement commands because the

measurement and math operations require several seconds to complete..

2.5 Send Acknowledge Command

The Send Acknowledge Command returns a simple status response which includes the address of

the sensor. Any measurement data in the sensor's buffer is not disturbed.

Command

Response

"a!"

"a"

Where:

a

Is the sensor address ("0-9", "A-Z", "a-z", "*", "?").