1-2 H-340 User manual

H-340

Figure 1 Bucket tip mechanism sensitivity

1.3 Installation

Install the H-340 in an open location where it will not be in the rain shadow of trees or buildings.

A rooftop location helps prevent vandalism and the accumulation of leaves and debris. Bolt or

screw the 3-mounting feet to the structure to prevent upset in high winds. Adjust the three self-

leveling screws until the “bulls eye” level is centered.

1.4 Factory Calibration

The H-340 is calibrated in the factory with an apparatus which provides a constant 4in/hour flow

rate. The water exiting from the bottom of the rain gauge is captured in a precision graduated

cylinder. After 100 bucket tips or more have occurred, the flow is stopped and the volume in the

graduated cylinder is measured. The procedure is as follows: