Wear parts and accessories – YSI IoLine Micro – 3 in 1 User Manual
Page 9
T 814245
04/2014
9
IoLine
Wear parts and accessories
Cleaning
Remove water-soluble contamination by rinsing with deionized water. Re-
move other contamination as follows:
Contamination
Cleaning procedure
Fat and oil
Rinse with water containing household
washing-up liquid
Lime and hydroxide deposits Rinse with citric acid (10 % by weight)
Proteins
Immerse in pepsin cleaning solution PEP/pH
for approx. 1 hour.
Note: Make sure the level of the reference
electrolyte is above that of the cleaning
solution.
Note
Hydrofluoric acid, hot phosphoric acid and strong alkaline solutions destroy
the glass membrane.
After cleaning
Rinse the combination electrode with deionized water and condition it in
reference electrolyte solution (KCl 3 mol/l, Ag
+
free) for at least 1 hour. Then
recalibrate the combination electrode.
Wear parts and accessories
Description
Model
Order no.
Reference electrolyte solution 250 ml
(KCl 3 mol/l, Ag
+
free)
KCl-250
109 705Y