2 cleaning, 3 what to do if, 2cleaning – YSI MultiLab IDS 4320 Conductivity Probe User Manual
Page 4: 3what to do if
Cleaning
4320
4
ba76140e01
02/2013
2
Cleaning
CAUTION
To clean the sensor, disconnect it from the instrument.
Outside cleaning
We recommend to clean the sensor thoroughly, especially before measuring
low conductivity values.
After cleaning, thoroughly rinse with deionized water and recalibrate if
necessary.
Aging of the
conductivity
measuring cell
Normally, the conductivity measuring cell does not age. Special measuring
media (e.g. strong acids and bases, organic solvents) or temperatures that
are too high may considerably reduce its lifetime or lead to damage. The
warranty does not cover failure caused by measuring conditions and
mechanical damage.
Disposal
We recommend to dispose of the measuring cell as electronic waste.
3
What to do if...
Contamination
Cleaning procedure
Lime sediments
Immerse in acetic acid for 5 minutes (volume
share = 10 %)
Fat/oil
Clean with warm water that contains
washing-up liquid
Error symptom
Cause
Remedy
No temperature or
conductivity display
– No connection between meter and
conductivity measuring cell
– Cable defective
– Establish connection between
meter and conductivity measuring
cell
Measurement delivers
implausible conductivity
values
– Measuring range exceeded
– Contamination in the area of the
electrodes
– Electrodes damaged
– Make sure the correct sensor is
being used for the application
– Clean the conductivity measuring
cell (see section 2).
– Return the sensor