4210
Commissioning, measuring, checking
5
ba76138e01
02/2013
At the same time, the level of the reference electrolyte must be at least 2 
cm above the level of the solution.
Conversion to
normal hydrogen
electrode
U
H
= U
Meas
+ U
Ref
U
Ref
is temperature dependent and can be taken from the following table (see
also DIN 38404-6):
4210
with:
U
H
= ORP, referring to the normal hydrogen electrode
U
Meas
= Measured ORP
U
Ref
= Voltage of the reference system compared to the normal
hydrogen electrode
T (°C)
T (°F)
U
Ref
[mV]
T (°C)
T (°F)
U
Ref
[mV]
4210
4210
0
32
+224
35
95
+200
5
41
+221
40
104
+196
10
50
+217
45
113
+192
15
59
+214
50
122
+188
20
68
+211
55
131
+184
25
77
+207
60
140
+180
30
86
+203
                    See also other documents in the category YSI Equipment:                                                            
                    - 3100 (50 pages)
 - 3200 (86 pages)
 - 5000 (70 pages)
 - 5200 (148 pages)
 - 5200A (128 pages)
 - 5200A (121 pages)
 - 5400 (127 pages)
 - 5239 Dissolved Oxygen Probe (23 pages)
 - 55 (26 pages)
 - 550A (18 pages)
 - 556 MPS - Quick Start Guide (4 pages)
 - 58 (40 pages)
 - 5905 (22 pages)
 - 6600V2 - User Manual (379 pages)
 - 60 (33 pages)
 - 63 (46 pages)
 - 650 MDS (72 pages)
 - 9300 Photometer (11 pages)
 - 9500 Photometer (143 pages)
 - 9500 Photometer (18 pages)
 - CR 3200 Thermoreactor (42 pages)
 - CR 4200 Thermoreactor (42 pages)
 - CR2200 Thermoreactor (36 pages)
 - DO200 (40 pages)
 - EC300 (40 pages)
 - EC300A (40 pages)
 - EC30A (12 pages)
 - IFL 700 IQ (46 pages)
 - IQ SensorNet 2020 XT Terminal (196 pages)
 - IQ SensorNet FDO Sensor (48 pages)
 - IQ SensorNet MIQ/A-MOD Module (42 pages)
 - IQ SensorNet MIQ/A-PR Module (36 pages)
 - IQ SensorNet MIQ/IC2 Module (30 pages)
 - IQ SensorNet MIQ/IF232 Module (18 pages)
 - IQ SensorNet MIQ/JB Module (18 pages)
 - IQ SensorNet MIQ/JBR Module (20 pages)
 - IQ SensorNet MIQ/PS Module (22 pages)
 - IQ SensorNet MIQ/R6 Module (56 pages)
 - IQ SensorNet NitraLyt Sensor (64 pages)
 - IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
 - IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
 - IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
 - IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
 - MultiLab 4010-1 (108 pages)
 - MultiLab 4010-2 (138 pages)
 