Technical data – YSI SensoLyt serries User Manual
Page 6
Technical data
SensoLyt
®
combination electrodes
6
ba76005e01
01/2012
Technical data
pH measuring
range
SensoLyt
®
SEA
SensoLyt
®
ECA
pH 2 ... 12
SensoLyt
®
SEA-HP
pH 4 ... 12
SensoLyt
®
DWA
pH 0 ... 14
ORP measuring
range
SensoLyt
®
PtA
-2000 ... 2000 mV
(pH application range pH 4 ... 12)
Pressure range at
temperature
(armored
electrodes only)
SensoLyt
®
SEA
SensoLyt
®
ECA
SensoLyt
®
DWA
SensoLyt
®
PtA
Temperature
Allowed overpressure
0 °C (32 °F)
1000 kPa (10 bar)
20 °C (68 °F)
1000 kPa (10 bar)
30 °C (86 °F)
500 kPa (5 bar)
40 °C (104 °F)
300 kPa (3 bar)
60 °C (140 °F)
100 kPa (1 bar)
SensoLyt
®
SEA-HP
0-60 °C (32-140 °F)
1000 kPa (10 bar)
All SensoLyt
®
combination electrodes meet the requirements of article 3 (3)
of the directive 97/23/EC ("Pressure equipment directive").
Measuring
electrode
SensoLyt
®
PtA
Platinum ring
All except for
SensoLyt
®
PtA
Glass
Reference
electrolyte,
junction
SensoLyt
®
SEA
SensoLyt
®
SEA-HP
SensoLyt
®
PtA
Gel polymer solid electrolyte;
2-hole junction
SensoLyt
®
ECA
Gel electrolyte; 1-hole junction
SensoLyt
®
DWA
Modified gel electrolyte;
ceramic junction
Shunt conduction
element
Ag/AgCl
Connection
Plug-in system (S7)
Minimum
immersion depth
25 mm