Specifications, Recommended spare parts list – YSI EC300A User Manual
Page 10

8
SPECIFICATIONS
Display Range
Accuracy
Resolution
Conductivity,
Auto-ranging
0.0 to 499.9 uS/cm
500 to 4999 uS/cm
5.00 to 49.99 mS/cm
50.0 to 200.0 mS/cm
±1% of reading plus 2 uS/cm
±1% of reading plus 5 uS/cm
±1% of reading plus 0.05 uS/cm
±2.5% of reading plus 0.5 mS/cm
0.01 uS/cm
1 mS/cm
0.01 mS/cm
0.1 mS/cm
Salinity
0.0 to 70.0 ppt
0.2% Full Scale
0.1 ppt
Temperature
°C
-10.0 to 90 °C
±0.2 °C or ±0.4% Full Scale,
whichever is greater
0.1 °C
Reference Temperature
15.0 to 25.0
°C
Temperature Coefficient
0.0% to 4.0%
Cell Constant
5.00
±
0.50
TDS Constant Range
0.30 to 1.00
Power/Battery life
One 9V battery
Approximately 500 hour
Calibration Back-up
Yes
Audio Feedback
Yes, on all touch keys
Instrument Case
Waterproof when connector cap installed, IP 67
Operating Temp. Range
0 to 50
°C
Operating Relative Humidity Range
up to 95%
Temperature Probe
Thermistor, 10k / 25
°C
Dimensions (L x W x H)
18.7 cm x 7.6 cm x 3.8 cm (7.37 in x 3 in x 1.5 in)
Weight (batteries included)
270 grams (.6 lb)
RECOMMENDED SPARE PARTS LIST
PART #
DESCRIPTION
300-1
1-meter probe and cable assembly
300-4
4-meter probe and cable assembly.
300-10
10-meter probe and cable assembly.
606043
Carrying case, hard sided.
485
Carrying case, soft sided.
Item #606042REF
Revision A, July 2012