Commissioning
CR 3200
12
ba76120e01
08/2012
3.2
Initial commissioning
Note
The thermoreactor works at an ambient temperature of +5 °C to
+40 °C. When the thermoreactor was transported from a cold
environment to a warm environment, condensate may occur and cause
a malfunction. Wait until the thermoreactor has adjusted to the new
environmental conditions before putting it into operation again (see
also chapter 8 T
).
Setting up the
thermoreactor
Adjusting the mains
voltage
1
Place the thermoreactor firmly onto a heat-resistant
underground.
2
Make sure that there is enough space between the
thermoreactor and other instruments or devices that are heat-
sensitive.
3
Check whether the arrow on the housing points to the mains
voltage (115 or 230 V) given on the fuse holder (3) that is
provided by the mains.
4
If the wrong mains voltage is set, perform steps 5 to 7.
5
Pull out the fuse holder (3).
6
Turn the fuse holder (3) so that the arrow on the housing points
to the mains voltage (115 or 230 V) provided by the mains.
7
Push the fuse holder (3) in completely.
2
3
1