11
S
ERVICING
The 5905/5010/200-BOD/Pro-BOD has four user replaceable
parts. The motor assembly can be replaced if it becomes noisy,
unbalanced, or malfunctions. The probe assembly, housing
assembly, and stir paddle can also be replaced if they become
damaged.
To disassemble the probe, perform the following steps while
referring to figure 3:
1. Remove stir paddle by pulling it straight out.
2. Remove the two screws (A) on the outside of the housing
using a 3/32 Allen wrench.
3. Loosen the top nut of the cable strain relief (B) from the
housing. This will allow the cable to feed through the housing
assembly.
4. Pull the probe and motor assembly out of the housing and
disconnect the two plugs (C).
5. To remove the motor assembly from the probe assembly,
remove the two screws (D) using the 3/32 Allen wrench.