2 quality control certification, Advanced instruments inc – Analytical Industries GPR-35 Oxygen Analyzer User Manual
Page 3

Advanced Instruments Inc.
3
2 Quality Control Certification
Date:
Customer: Order No.:
Pass
Model:
GPR-35 Oxygen Analyzer Ambient Area Monitoring S/N _______________
Sensor:
( ) GPR-11-32-RTS Oxygen Sensor
( ) XLT-11-24-RTS Oxygen Sensor
( ) Remote above sensor -4R suffix S/N _______________
Approvals:
CSA/NRTL and CE
Accessories:
Owner’s Manual
A-2344 Calibration Flow Through Adapter
Configuration:
A-1113-C PCB Assembly Main/Display
A-1114-C PCB Assembly Power Supply
Range: 0-25%
Low and Low, Low Alarms: 20.0% CAUTION, 19.5% DANGER
Test – AC Power
Calibrates at 20.9% oxygen in ambient air with adequate span
CAUTION/LOW O2 Alarms: LED green–safe/red–alarm, relay contact
DANGER/LOW, LOW O2 Alarms: LED green–safe/red–alarm, relay contact
Alarm relays activate/deactivate with changes in O
2
concentration
LED indicators: AC FAIL-red, LOW BATT-red
Sensor failure 5V +0.5V
Power failure relay alarm contacts
Analog signal output 0-1V
Test – Battery
No interruption in operation when disconnecting AC power
Calibrates at 20.9% oxygen in ambient air with adequate span
CAUTION/LOW O2 Alarms: LED green–safe/red–alarm, relay contact
DANGER/LOW, LOW O2 Alarms: LED green–safe/red–alarm, relay contact
Alarm relays activate/deactivate with changes in O
2
concentration
LED indicators: AC FAIL-red, LOW BATT-red
Final
Overall inspection for physical defects
Options:
( ) High, Low Alarms: HIGH O2 23%, LOW O2 19.5%
( ) 4-20mA isolated signal output (0-1VDC negative ground standard)
( ) A-2501 Integral sampling pump
( ) Integral audible alarm
( ) Remote sensor: A-1114C-R PCB Assy; components see following page
Notes: