beautypg.com

Calibration & accuracy – Analytical Industries GPR-1000 ATEX Portable Trace PPM Oxygen Analyzer User Manual

Page 9

background image



9

Calibration & Accuracy

Single Point Calibration: As previously described the galvanic oxygen sensor generates an electrical current

sensor exhibiting an absolute zero, e.g. the sensor does not generate a current output in the absence of oxygen.

Given these linearity and absolute zero properties, single point calibration is possible.
Pressure: Because sensors are sensitive to the partial pressure of oxygen in the sample gas their output is a
function of the number of molecules of oxygen 'per unit volume'. Readouts in percent are permissible only when

the total pressure of the sample gas being analyzed remains constant. The pressure of the sample gas and that of

the calibration gas(es) must be the same (reality < 1-2 psi).
Temperature: The rate oxygen molecules diffuse into the sensor is controlled by a Teflon membrane otherwise

known as an 'oxygen diffusion limiting

barrier' and all diffusion processes

are

temperature sensitive, the fact the

sensor's electrical output will vary

with

temperature is normal. This variation

is

relatively constant 2.5% per ºC. A

temperature compensation circuit

employing a thermistor offsets this
effect with an accuracy of +5% or

better and generates an output

function that is independent of

temperature. There is no error if the

calibration and sampling are

performed at the same temperature

or if

the measurement is made immediately after calibration. Lastly, small temperature variations of 10-15º produce <

+1% error.
Accuracy:

In light of the above parameters, the overall accuracy of an analyzer is affected by two types of

errors: 1) those producing 'percent of reading errors', illustrated by Graph A below, such as +5% temperature

compensation

circuit, tolerances of range resistors and the 'play' in the potentiometer used to make span

adjustments and 2) those producing 'percent of full scale errors', illustrated by Graph B, such as +1-2% linearity
errors in readout devices, which are really minimal due to today's technology and the fact that other errors are

'spanned out' during calibration.
Graph C illustrates these 'worse case' specifications that are typically used to develop an analyzer's overall accuracy

statement of +2% of full scale at constant temperature or +5% over the operating temperature range. QC testing

is typically <+0.5% prior to shipment.







This manual is related to the following products: