beautypg.com

8 troubleshooting, Symptom possible cause recommended action – Analytical Industries GPR-1000 ATEX Portable Trace PPM Oxygen Analyzer User Manual

Page 26

background image



26

8 Troubleshooting

Symptom

Possible Cause

Recommended Action

Slow recovery



At installation, defective sensor


Air leak in sample system connection(s)


Abnormality in zero gas

Damaged in service - prolonged

exposure to air, electrolyte leak

Sensor nearing end of life

Replace sensor if recovery unacceptable or

O

2

reading fails to reach 10% of lowest

range


Leak test the entire sample system:

Vary the flow rate, if the O

2

reading

changes inversely with the change in flow

rate indicates an air leak - correct source

of leak


Qualify zero gas (using portable analyzer)

Replace sensor

Replace sensor

High O

2

reading

after installing or

replacing sensor

Analyzer calibrated before sensor

stabilized caused by:

1) Prolonged exposure to ambient air,

worse if sensor was unshorted


2) Air leak in sample system

connection(s)

3) Abnormality in zero gas

Allow O

2

reading to stabilize before making

the span/calibration adjustment

Continue purge with zero gas


Leak test the entire sample system

(above)

Qualify zero gas (using portable analyzer)

High O

2

reading

Sampling

Flow rate exceeds limits

Pressurized sensor


Improper sensor selection

Abnormality in gas

Correct pressure and flow rate

Remove restriction on vent line or open

SHUT OFF valve completely

Replace GPR/PSR sensor with XLT sensor

when CO

2

or acid gases are present

Qualify the gas (use a portable analyzer)

Response time

slow

Air leak, dead legs, distance of sample

line, low flow rate, volume of optional

filters and scrubbers

Leak test (above), reduce dead volume or

increase flow rate

This manual is related to the following products: