6. MAINTENANCE
1) 9-Volt battery replacement
a. Ensure the instrument is not connected to any external circuit. Set
the selector switch to OFF position and remove the test leads from
terminals.
b. Remove the screw on the bottom case and lift the bottom case.
c. Remove the spent battery and replace it with a battery of the same
type.
2) Fuse replacement
a. Ensure the instrument is not connected to any external circuit. Set
the selector switch to OFF position and remove the test leads from
terminals.
b. Remove the screw on the bottom case and lift the bottom case.
c. Replace the fuse with the same type and rating: 5×20mm,
100mA/250V, fast-blow fuse or as the replacements.
The specifications are subject to change without notice.