M6 Specifications
Description
Specification
Test capacity
2-6 valve samples
Valve test chamber
6; design based on customer requirements
Valve test chamber material
Acrylic, for full visibility of valve status at all times
Sample type
Tissue and Mechanical Heart Valves, Septal Defect Closures, Artificial
Heart Bladders, Membranes/sheets, Sutures/clips; Intra-aortic balloons,
Percutaneously-delivered Valves
Motor
¼ hp DC (½ hp DC optional)
Volume capacity
1-2 liters
Measured parameters
Cycle count, frequency, pressure, temperature
Fluid temperature
25º to 40ºC
Temperature accuracy
±0.3ºC of set temperature
Speed
Up to 33Hz (50Hz with the optional ½ hp motor*)
Speed control
Closed loop feedback type, user adjustable
Temperature control
Microprocessor based with RTD sensor
Working fluid
PBS, water, blood analog
Pressure transducers
Two optional; one each for inflow and outflow
Back pressure on test valve
User adjustable via Throttle Valve
Pump type
Pistonless Swash Plate, adjustable, with bellows
Controller
Controller features digital speed count,
pressure and temperature readouts, and drive controls
Software
Invitroflow software package, with intuitive user interface to monitor
and
control
the
M6
Capacitance tank
2 liters capacity, to provide air-water interface between the Controller
and
the
M6
Dimensions
12” Dia. X 22.5” H
Power requirement
115 AC
Weight
72
lbs,
dry
Specifications are subject to change without notice.
*Sample design, size and compliance will determine the highest frequency at which the test can be performed with high confidence.
Dynatek’s Model M6 Heart Valve Tester Specifications
Optional table for M6, with manual tilt top,
to assist in debubbling operations
M6™ and Invitroflow™ are trademarks of Dynatek Labs, Inc.