Specifications – Velleman DVM851 User Manual
Page 6

DVM851
01.08.2012
6
condensation is to be expected.(home and office environments fall under this
category)
Pollution degree 3 Conductive pollution occurs, or dry nonconductive pollution occurs that becomes
conductive due to condensation that is to be expected.
(industrial environments and environments exposed to outside air
contact with precipitation)
Pollution degree 4 The pollution generates persistent conductivity caused by conductive dust or by rain
or snow. (exposed outdoor environments and environments where high humidity
levels or high concentrations of fine particles occur)
Warning:
This device was designed in accordance with EN 61010-1 pollution degree 2. This implies that certain
restrictions in use apply that are related to pollution which can occur within the environment of use.
Refer to the table above.
This device is only suitable for measurements in Pollution degree class 2 environments.
9.
Specifications
This device is not calibrated when purchased!
Regulations concerning environment of use:
•
Use this meter only for measurements in CAT I and CAT II environments (see §7)
•
Use this meter only in a pollution degree 2 environment (see §8)
Ideal working conditions include:
•
Temperature: 18 to 28°C (64°F to 82°F)
•
relative humidity: max. 80%
•
altitude: max. 2000m (6560ft)
voltage
max. 600V
Fuse protection
F 500mA/250V, F10A/250V
Power
9V alkaline battery 6LR61
Display
LCD, 1999 counts
sampling rate
2-3/sec.
Measuring method
Dual-slope integration A/D converter
Overrange indication
Only figure "1" on the display
Polarity indication
"-" displayed for negative polarity
Operating temperature
0 to 40°C
Storage temperature
-10°C to 50°C
Low battery indication
appears on the display
Dimensions
138 x 69 x 31mm
Weight
Approx. 142g
9.1 DC VOLTAGE
Do not measure circuits that may contain voltages > 600V
Range
Resolution
Accuracy
200mV
100µV
±
0.5% of rdg
2V
1mV
±
0.8% of rdg
20V
10mV
200V
100mV
600V
1V
±
1.0% of rdg
9.2 DC CURRENT
Do not measure current in circuits with voltages > 250V
Range
Resolution
Accuracy
200µA
0.1µA
±
1% of rdg
2mA
1µA
20mA
10µA
200mA
100µA
±
1.5% of rdg
10A
10mA
±
3% of rdg
Overload protection by fuse: 200mA range F 500mA/250V, 10A range F10A/250V.
©Velleman nv
condensation is to be expected.(home and office environments fall under this
tive pollution occurs, or dry nonconductive pollution occurs that becomes
(industrial environments and environments exposed to outside air - but not in
he pollution generates persistent conductivity caused by conductive dust or by rain
or snow. (exposed outdoor environments and environments where high humidity
. This implies that certain
restrictions in use apply that are related to pollution which can occur within the environment of use.
lution degree class 2 environments.
Use this meter only for measurements in CAT I and CAT II environments (see §7)
10A/250V
6LR61
slope integration A/D converter
Only figure "1" on the display
" displayed for negative polarity
appears on the display
Accuracy
0.5% of rdg
±
2 digits
0.8% of rdg
±
2 digits
1.0% of rdg
±
2 digits
Accuracy
1% of rdg
±
2 digits
1.5% of rdg
±
2 digits
3% of rdg
±
2 digits
250V.