beautypg.com

6 supplement, 1 technical data – VEGA VEGABAR 83 4 … 20 mA - Quick setup guide User Manual

Page 15

background image

15

6 Supplement

VEGABAR 83 • 4 … 20 mA

46312-EN-131125

6 Supplement

6.1 Technical data

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry

Ʋ Cable gland

M20 x 1.5

Ʋ Cable diameter (options)

5 … 9 mm (0.20 … 0.35 in)
6 … 12 mm (0.24 … 0.47 in)
10 … 14 mm (0.39 … 0.55 in)

Ʋ Cable entry

½ NPT

Ʋ Blind plug

M20 x 1.5; ½ NPT

Ʋ Closing cap

M20 x 1.5; ½ NPT

Plug options

Ʋ Signal circuit

Plug M12 x 1, according to ISO 4400, Harting HAN,

7/8" FF

Ʋ Indication circuit

M12 x 1 plug

Wire cross-section (spring-loaded terminals)

Ʋ Massive wire, cord

0.2 … 2.5 mm² (AWG 24 … 14)

Ʋ Stranded wire with end sleeve

0.2 … 1.5 mm² (AWG 24 … 16)

Voltage supply
Operating voltage U

B

Ʋ Non-Ex instrument

9.6 … 35 V DC

Ʋ Ex-d instrument

9.6 … 35 V DC

Ʋ Ex-ia instrument

9.6 … 30 V DC

Ʋ Ex-d-ia instrument

15 … 35 V DC

Ʋ Ex-d-ia instrument with ship approval 15 … 35 V DC

Interpolation protection

Integrated

Permissible residual ripple - Non-Ex, Ex-ia instrument

Ʋ for U

N

12 V DC (9.6 V< U

B

< 14 V)

≤ 0.7 V

eff

(16 … 400 Hz)

Ʋ for U

N

24 V DC (18 V< U

B

< 35 V)

≤ 1.0 V

eff

(16 … 400 Hz)

Permissible residual ripple - Ex-d-ia instrument

Ʋ for U

N

24 V DC (18 V< U

B

< 35 V)

≤ 1 V

eff

(16 … 400 Hz)

Load resistor

Ʋ Calculation

(U

B

- U

min

)/0.022 A

Ʋ Example - Non-Ex instrument with

U

B

= 24 V DC

(24 V - 9.6 V)/0.022 A = 655 Ω