11

5 Set up with the display and adjustment module

VEGABAR 83 • 4 … 20 mA

46312-EN-131125

Carry out the following steps in the below sequence.
You can find "Extended adjustment" in the next sub-chapter.
1. Measurement loop name

In the first menu item you assign a suitable measurement loop name.

Permitted are names with max. 19 characters.
2. Application

In this menu item, you activate/deactivate the slave for the electronic

differential pressure and select the application. The application com-

prises process pressure and level measurement.
3. Units

In this menu item you determine the adjustment and temperature

units of the instrument. Depending on the selected application in the

menu item "Application", different adjustment units are available.

4. Position correction

In this menu item you compensate the influence of the installation

position of the instrument (offset) on the measured value.
5. Zero adjustment

In this menu item you carry out the zero adjustment for the processs

pressure.
Enter the respective pressure value for 0 %.
6. Span adjustment

In this menu item you carry out the span adjustment for the processs

pressure
Enter the respective pressure value for 100 %.

4. Position correction

In this menu item you compensate the influence of the installation

position of the instrument (offset) on the measured value.
5. Max. adjustment

In this menu item you carry out the max. adjustment for level
Enter the percentage value and the corresponding value for the max.

level.

Presettings

Quick setup - Process

pressure measurement

Quick setup - Level meas-

urement