3 Mounting

VEGABAR 87 • Modbus and Levelmaster protocol

46330-EN-131125

3 Mounting

3.1 General instructions for use of the instrument

Protect your instrument against moisture penetration through the fol-

lowing measures:

Use the recommended cable (see chapter "Connecting to power

supply")

Tighten the cable gland

Turn the housing in such a way that the cable gland points down-

ward

Loop the connection cable downward in front of the cable gland

This applies particularly to:

Outdoor mounting

Installations in areas where high humidity is expected (e.g. through

cleaning processes)

Installations on cooled or heated vessels

3.2 Ventilation and pressure compensation

The filter element is mounted into the electronics housing. It has the

following functions:

Ventilation of the electronics housing

Atmospheric pressure compensation (with relative pressure meas-

uring ranges)

Turn the housing so that the filter element points downward after

the instrument is installed. This provides better protection against

buildup.

Protection against mois-

ture

Instruments in non-Ex,

Ex-ia and Ex-d-ia version