beautypg.com

3 product description, 1 configuration – VEGA VEGAFLEX series 80 - External housing User Manual

Page 5

background image

5

3 Product description

External housing • VEGAFLEX 80 series

46802-EN-131120

3 Product description

3.1 Configuration

The scope of delivery encompasses:

Level sensor with external housing

High frequency connection cable (HF)

Documentation

– a device operating instructions manual

– This supplementary instructions manual

– Ex specific safety instructions (with Ex versions), if necessary

further certificates

The instrument version "Cable outlet IP 68, external housing" consists

of a sensor housing, the instrument housing and a confectioned HF

connection cable.

A

B

1

2

3
4

5

Fig. 1: Components of the external housing for plics

®

devices, N-plug

A Sensor housing

B Instrument housing

1 Screw-on cover

2 Base element

3 Housing base

4 Wall mounting plate

5 HF connection cable

The electrical connection must be carried out on the instrument hous-

ing.
The instrument housing consists of the base element, the screwed

cover for the electronics or connection compartment and the housing

socket.
Depending on the order, the screwed cover of the instrument housing

is available with or without inspection window for the display and

adjustment module PLICSCOM.
The HF connection cable of the external housing can influence the

measurement properties of VEGAFLEX. When the length of the

sensor exceeds the length of the HF connection cable, then the min.

dielectric constant is ε

r

> 2 independent of the instrument version. If

the HF connection cable is longer, there are no limitations.

Scope of delivery

Constituent parts