VEGA VEGAFLEX 86 4 … 20 mA_HART - four-wire Coax probe User Manual
Page 42
42
6 Set up with the display and adjustment module
VEGAFLEX 86 • 4 … 20 mA/HART - four-wire
42285-EN-130314
The following reset functions are available:
Delivery status: Restoring the parameter settings at the time of ship-
ment from the factory incl. the order-specific settings. A created false
signal suppression, user-programmable linearization curve as well as
the measured value memory will be deleted.
Basic settings: Resetting the parameter settings incl. special and
laboratory parameters to the default values of the respective instru-
ment. A created false signal suppression, user programmable lineari-
zation curve as well as the measured value memory will be deleted.
The following table shows the default values of the instrument. De-
pending on the instrument version or application, all menu items may
not be available or some may be differently assigned:
Menu
Menu item
Default value
Setup
Block operation
Released
Measurement
loop name
Sensor
Units
Distance unit: mm
Temperature unit: °C
Probe length
Length of the probe Ex factory
Product type
Liquid
Application
Level, vessel
Medium, dielectric
constant
Water-based, > 10
Max. adjustment
0.000 m(d) - note blocking distances
Min. adjustment
Probe length - take dead band into
account
Linearization
Linear
Damping
0.0 s
Scaling size
Volume in l
Current output,
adjustment
0.00 lin %, 0 l
100.00 lin %, 100 l
Current output
mode
Last measured value
Current output
Min./Max.
Min. current 3.8 mA, max. current
20.5 mA
Display
Language
As set in the device
Displayed value 1 Filling height Level
Displayed value 2 Electronics temperature
Backlight
Switched off