VEGA VEGABAR 83 Save sensor - Operating Instructions User Manual
Page 48
48
9 Supplement
VEGABAR 83 • Slave sensor for electronic differential pressure
45051-EN-131011
Ambient conditions
Version
Ambient temperature
Storage and transport temper-
ature
Standard version
-40 … +80 °C (-40 … +176 °F)
-60 … +80 °C (-76 … +176 °F)
Version IP 66/IP 68 (1 bar)
-20 … +80 °C (-4 … +176 °F)
-20 … +80 °C (-4 … +176 °F)
Version IP 68 (25 bar) with connec-
tion cable PUR
-20 … +80 °C (-4 … +176 °F)
-20 … +80 °C (-4 … +176 °F)
Version IP 68 (25 bar), connection
cable PE
-20 … +60 °C (-4 … +140 °F)
-20 … +60 °C (-4 … +140 °F)
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry
Ʋ Cable diameter (options)
5 … 9 mm (0.20 … 0.35 in)
6 … 12 mm (0.24 … 0.47 in)
10 … 14 mm (0.39 … 0.55 in)
Ʋ Cable entry
½ NPT
Ʋ Blind plug
M20 x 1.5; ½ NPT
Ʋ Closing cap
M20 x 1.5; ½ NPT
Plug options
Ʋ Signal circuit
Plug M12 x 1, according to ISO 4400, Harting HAN,
7/8" FF
Ʋ Indication circuit
M12 x 1 plug
Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, cord
0.2 … 2.5 mm² (AWG 24 … 14)
Ʋ Stranded wire with end sleeve
0.2 … 1.5 mm² (AWG 24 … 16)
Electromechanical data - version IP 68 (25 bar)
Connection cable between IP 68 instrument and external housing
Ʋ Configuration
four wires, one suspension wire, one breather capillary,
screen braiding, metal foil, mantle
Ʋ Wire cross-section
0.5 mm² (AWG 20)
Ʋ Wire resistance
< 0.036 Ω/m (0.011 Ω/ft)
Ʋ Standard length
5 m (16.40 ft)
Ʋ Max. length
180 m (590.5 ft)
Ʋ Min. bending radius at 25 °C/77 °F
25 mm (0.985 in)
Ʋ Diameter approx.
8 mm (0.315 in)
Ʋ Colour
Blue
Cable entry/plug
7)
7)
Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.