VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART four-wire Approval according to LPR radio standard User Manual
Page 71
Indication
LC display in dot matrix
Measured value indication
-
Number of digits
5
-
Size of digits
W x H = 7 x 13 mm (sensors with software ≥ 4.0.0,
hardware ≥ 2.0.0)
W x H = 5 x 8 mm (sensors with software ≤ 3.99,
hardware < 2.0.0)
Adjustment elements
4 keys
Protection rating
-
unassembled
IP
20
-
mounted into the sensor without cover
IP
40
Materials
-
Housing
ABS
-
Inspection window
Polyester foil
Integrated clock
Date format
Day.Month.Year
Time format
12 h/24 h
Time zone Ex factory
CET
Measurement electronics temerature
Resolution
1 °C (1.8 °F)
Accuracy
±
1 °C (1.8 °F)
Voltage supply
Operating voltage
-
Version for low voltage
9.6 … 48 V DC, 20 … 42 V AC, 50/60 Hz
-
Version for mains voltage
90 … 253 V AC, 50/60 Hz
Interpolation protection
Available
Max. power consumption
4 VA; 2.1 W
Electrical protective measures
Protection rating
IP
66/IP 67
Overvoltage category
III
Protection class
-
With mains voltage
I
-
With low voltage
II
Approvals
Instruments with approvals can have different technical data depending on the version.
VEGAPULS
62 • Approval according to LPR radio standard
71
11 Supplement
41719
-EN
-120301