beautypg.com

VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART four-wire Approval according to LPR radio standard User Manual

Page 71

background image

Indication

LC display in dot matrix

Measured value indication

-

Number of digits

5

-

Size of digits

W x H = 7 x 13 mm (sensors with software ≥ 4.0.0,
hardware ≥ 2.0.0)
W x H = 5 x 8 mm (sensors with software ≤ 3.99,
hardware < 2.0.0)

Adjustment elements

4 keys

Protection rating

-

unassembled

IP

20

-

mounted into the sensor without cover

IP

40

Materials

-

Housing

ABS

-

Inspection window

Polyester foil

Integrated clock

Date format

Day.Month.Year

Time format

12 h/24 h

Time zone Ex factory

CET

Measurement electronics temerature

Resolution

1 °C (1.8 °F)

Accuracy

±

1 °C (1.8 °F)

Voltage supply

Operating voltage

-

Version for low voltage

9.6 … 48 V DC, 20 … 42 V AC, 50/60 Hz

-

Version for mains voltage

90 … 253 V AC, 50/60 Hz

Interpolation protection

Available

Max. power consumption

4 VA; 2.1 W

Electrical protective measures

Protection rating

IP

66/IP 67

Overvoltage category

III

Protection class

-

With mains voltage

I

-

With low voltage

II

Approvals

Instruments with approvals can have different technical data depending on the version.

VEGAPULS

62 • Approval according to LPR radio standard

71

11 Supplement

41719

-EN

-120301