3 switch-on phase – VEGA VEGAPULS 68 (≥ 2.0.0 - ≥ 4.0.0) HART - integrated accumulator User Manual
Page 29
29
5 Connecting to power supply
VEGAPULS 68 • HART - integrated accumulator
40935-EN-121011
2
4
3
5
0
5
1
6
2
7
3
8
4
9
modes
4 ON/OFF
3 ON/OFF 1h
1,2 OFF
Charge
Accu
5..9 auto cycle
1
2
+
( )
(-)
1
Fig. 29: Supply room
1 Internal connection of the socket for the battery charger
2 Mode switch
3 LED green, charging process
4 LED yellow, charging status
5 Socket for the battery charger
The mode switch enables the selection of the following modes:
•
0 = sensor off, LEDs show the accumulator status
•
1, 2 = sensor off, LEDs off
•
3 = sensor on for 1 hour after pressing a key (delivery status)
•
4 = sensor permanently on, switching on/off via button
•
5 = sensor is switched on every 30 min. for 3 min.
•
6 = sensor is switched on every hour for 3 min.
•
7 = sensor is switched on every 6 hours for 3 min.
•
8 = sensor is switched on every 12 hours for 3 min.
•
9 = sensor is switched on every 24 hours for 3 min.
The green LED characterizes the charging process:
•
LED flashes = Accumulator is charging
•
LED lights = accumulator is full, battery charger should be
unplugged (accumulator life time)
After pressing the key or changing the mode the yellow LED shows
the accumulator status for approximately 10 s as follows:
•
LED lights = accumulator is full
•
LED flashes = accumulator should be charged
•
LED off = accumulator is empty
5.3 Switch-on phase
The instrument is switched on and off by means of a button outside
on the housing.
Supply room