beautypg.com

VEGA VEGABAR 51 4 … 20 mA User Manual

Page 12

background image

12

4 Mounting

VEGABAR 51 • 4 … 20 mA

36712-EN-130321

2

1

2

1

2

1

1

2

Fig. 4: Position of the filter element
1 Filter element

2 Blind plug

Caution:

Due to the filter effect, the pressure compensation is time delayed.

When opening/closing the housing cover quickly, the measured value

can change for a period of approx. 5 s by up to 15 mbar.
Information:

Make sure that the filter element is always free of buildup during

operation. A high-pressure cleaner may not be used for cleaning.
With instrument versions in protection IP 66/IP 68, 1 bar, the ventila-

tion is realised via the capillaries in the permanently connected cable.

The filter element is replaced by a blind plug.

Higher process temperatures often mean also higher ambient

temperatures. Make sure that the upper temperature limits stated in

chapter "Technical data" for the environment of the electronics hous-

ing and connection cable are not exceeded.

1

2

Fig. 5: Temperature ranges
1 Process temperature

2 Ambient temperature

Temperature limits