VEGA VEGABAR 67 4 … 20 mA_HART - climate compensated User Manual
Page 55
55
10 Supplement
VEGABAR 67 • 4 ... 20 mA/HART - climate compensated
39444-EN-130321
Weight approx.
Ʋ Basic weight
0.7 kg (1.543 lbs)
Ʋ Suspension cable
0.1 kg/m (0.07 lbs/ft)
Ʋ Connection tube
1.5 kg/m (1 lbs/ft)
Ʋ Straining clamp
0.2 kg (0.441 lbs)
Ʋ Threaded fitting
0.4 kg (0.882 lbs)
Lengths
Ʋ Connection tube
0.25 … 6 m (0.82 … 19.69 ft)
Output variable
Output signal
4 … 20 mA/HART
HART output values according to HART standard 5.0
Ʋ Primary Value
Pressure
Ʋ Secondary Value
Temperature
10)
Signal resolution
1.6 µA
Failure signal current output (adjustable) mA-value unchanged 20.5 mA, 22 mA, < 3.6 mA
Max. output current
22 mA
Load
see load diagram under Power supply
Met NAMUR recommendation
NE 43
Dynamic behaviour output
Run-up time approx.
10 s
90 %
100 %
10 %
t
t
T
A
t
t
S
2
1
Fig. 32: Sudden change of the process variable. t
T
: dead time; t
A
: rise time; t
S
: jump response time
1 Process variable
2 Output signal
Dead time
≤ 150 ms
Rise time
≤ 100 ms (10 … 90 %)
Step response time
≤ 250 ms (ti: 0 s, 10 … 90 %)
Damping (63 % of the input variable)
0 … 999 s, adjustable
10)
Depending on the respective instrument and the process fitting, the temperature can deviate from the real
process temperature.