10 Supplement

VEGABAR 52 • Foundation Fieldbus - climate compensated

36789-EN-130321

Measuring cell seal

Product temperature - standard

version

Product temperature - version

with extended temperature range

FFKM (Perlast G75B)

-15 … +120 °C (-4 … +248 °F)

-15 … +150 °C (5 … +302 °F)

Vibration resistance

mechanical vibrations with 4 g and 5 … 100 Hz

13)

Shock resistance

Acceleration 100 g/6 ms

14)

Electromechanical data - version IP 66/IP 67
Cable entry/plug

15)

Ʋ Single chamber housing

– 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT

or:

– 1␣x plug (depending on the version), 1␣x blind stopper

M20␣x␣1.5

or:

– 2 x blind plug M20 x 1,5

Ʋ Double chamber housing

– 1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5; plug M12 x 1 for the external

display and adjustment unit (optional)

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT, plug

M12 x 1 for the external display and adjustment unit

(optional)

or:

– 1 x plug (depending on the version), 1 x blind plug

M20 x 1.5; plug M12 x 1 for the external display and

adjustment unit (optional)

or:

– 2 x blind stoppers M20 x 1.5; plug M12 x 1 for the

external display and adjustment unit (optional)

Spring-loaded terminals for wire cross-

section

< 2.5 mm² (AWG 14)

Electromechanical data - version IP 66/IP 68 (1 bar)
Cable entry

Ʋ Single chamber housing

– 1 x IP 68 cable gland M20 x 1.5; 1 x blind plug

M20 x 1.5

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT

Connection cable

Ʋ Configuration

four wires, one suspension wire, one breather capillary,

screen braiding, metal foil, mantle

Ʋ Wire cross-section

0.5 mm² (AWG 20)

13)

Tested according to the guidelines of German Lloyd, GL directive 2.

14)

Tested according to EN 60068-2-27.

15)

Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.