beautypg.com

VEGA VEGABAR 54 4 … 20 mA_HART User Manual

Page 56

background image

56

10 Supplement

VEGABAR 54 • 4 … 20 mA/HART

36726-EN-130321

Protection rating

Ʋ unassembled

IP 20

Ʋ mounted into the sensor without cover IP 40

Material

Ʋ Housing

ABS

Ʋ Inspection window

Polyester foil

Voltage supply
Operating voltage

Ʋ Non-Ex instrument

12 … 36 V DC

Ʋ EEx-ia instrument

12 … 30 V DC

Operating voltage with illuminated display and adjustment module

Ʋ Non-Ex instrument

20 … 36 V DC

Ʋ EEx-ia instrument

20 … 30 V DC

Permissible residual ripple

Ʋ < 100 Hz

U

ss

< 1 V

Ʋ 100 Hz … 10 kHz

U

ss

< 10 mV

Load

see diagram

1000

750

500

250

12

18

16

14

20

22

24

26

28

30

32

34

36

V

4

1

2

3

Fig. 26: Voltage diagram
1 HART load with HART version

2 Voltage limit EEx-ia instrument

3 Voltage limit non-Ex instrument

4 Operating voltage

Electrical protective measures
Protection rating

Ʋ Housing, standard

IP 66/IP 67

16)

Ʋ Aluminium and stainless housing

(optionally available)

IP 68 (1 bar)

17)

Ʋ Process component in IP 68 version IP 68 (25 bar)

Ʋ External housing

IP 65, IP 66/IP 68 (0.2 bar)

16)

Instruments with gauge pressure measuring ranges cannot detect the ambient pressure when submerged, e.g.

in water. This can lead to falsification of the measured value.

17)

Only with instruments with absolute pressure ranges.