VEGA VEGAPULS 66 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART - two-wire User Manual
Page 66
66
11 Supplement
VEGAPULS 66 • 4 … 20 mA/HART - two-wire
36519-EN-121011
Ʋ Nominal pressure stage
PN 250
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry
Ʋ Cable gland
M20 x 1.5 (cable: ø 5 … 9 mm)
Ʋ Cable entry
½ NPT
Ʋ Blind stopper
M20 x 1.5; ½ NPT
Ʋ Closing cap
M20 x 1.5; ½ NPT
Plug options
Ʋ Signal circuit
M12 x 1, according to ISO 4400, Harting HAN, 7/8" FF
Ʋ Indication circuit
M12 x 1
Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, cord
0.2 … 2.5 mm² (AWG 24 … 14)
Ʋ Stranded wire with end sleeve
0.2 … 1.5 mm² (AWG 24 … 16)
Electromechanical data - version IP 66/IP 68 (1 bar)
Options of the cable entry
Ʋ Cable gland with integrated connec-
tion cable
M20 x 1.5 (cable: ø 5 … 9 mm)
Ʋ Cable entry
½ NPT
Ʋ Blind stopper
M20 x 1.5; ½ NPT
Connection cable
Ʋ Wire cross-section
0.5 mm² (AWG 20)
Ʋ Wire resistance
< 0.036 Ω/m
Ʋ Tensile strength
< 1200 N (270 lbf)
Ʋ Standard length
5 m (16.4 ft)
Ʋ Max. length
180 m (590.6 ft)
Ʋ Min. bending radius
25 mm (0.984 in) with 25 °C (77 °F)
Ʋ Diameter approx.
8 mm (0.315 in)
Ʋ Colour - Non-Ex version
Blue
Ʋ Colour - Ex-version
Blue
Indicating and adjustment module
Voltage supply and data transmission
through the sensor
Indication
LC display in dot matrix
Measured value indication
Ʋ Number of digits
5
Ʋ Size of digits
W x H = 7 x 13 mm
Adjustment elements
4 keys
Protection rating
Ʋ unassembled
IP 20
Ʋ mounted into the sensor without cover IP 40