Quick start – VEGA VEGAPULS 66 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART - two-wire User Manual
Page 2
2
Quick start
VEGAPULS 66 • 4 … 20 mA/HART - two-wire
36519-EN-121011
Quick start
The quick start enables a quick setup with many applications. You
can find further information in the respective chapters of the operating
instructions manual.
1. Distance from the vessel wall > 200 mm, the antenna should
protrude > 10 mm into the vessel
> 500 mm
(19.69")
ca. 10 mm
Fig. 1: Distance of the antenna to the vessel wall/vessel ceiling
2. Note min. socket diameter depending on the socket length
For further information see chapter "Mounting".
1. Make sure that the power supply corresponds to the specifica-
tions on the type label.
2. Connect the instrument according to the following illustration
5
1
2
+
( )
(-)
6 7 8
4...20mA
2
3
4
1
Fig. 2: Electronics and connection compartment, single chamber housing
1 Voltage supply, signal output
2 For indicating and adjustment module or interface adapter
3 For external indicating and adjustment unit
4 Ground terminal for connection of the cable screen
For further information see chapter "Connecting to power supply".
1. Go via the indicating and adjustment module to the menu "Setup".
2. In the menu item "Medium"you select the medium of your applica-
tion, for example "Aqueous solution".
Mounting
Connect electrically
Set parameters