VEGA VEGABAR 52 Profibus PA - climate compensated User Manual
Page 53
53
10 Supplement
VEGABAR 52 • Profibus PA - climate compensated
36788-EN-130321
Ʋ Version IP 66/IP 68 (1 bar), connec-
tion cable PE
-20 … +60 °C (-4 … +140 °F)
Process conditions
The specifications of the pressure stage and product temperature are used as an overview. The
specifications on the type label are applicable.
Pressure stage, process fitting
Ʋ Thread 316L, depending on connec-
tion
PN 10, PN 60, PN 160
Ʋ Thread Aluminium
PN 25
Ʋ Thread PVDF
PN 10
Ʋ Hygienic fittings 316L, depending on
connection
PN 6, PN 10, PN 25, PN 40 (PN 40 only with DRD and
DIN 11851)
Ʋ Flange 316L
PN 16, PN 40,150 lbs, 300 lbs, 600 lbs
Ʋ Flange with extension 316L
without PN specification, PN 16, PN 40 or 150 lbs,
300 lbs, 600 lbs
Ʋ Flange flattened on both sides 316L
PN 10
Ʋ Flange PVDF
PN 16
Product temperature depending on the measuring cell seal
12)
Measuring cell seal
Product temperature - standard
version
Product temperature - version
with extended temperature range
FKM (VP2/A)
-20 … +120 °C (-4 … +248 °F)
-20 … +150 °C (-4 … +302 °F)
FKM (A+P 70.16)
-40 … +120 °C (-40 … +248 °F)
-
EPDM (A+P 75.5/KW75F)
-40 … +120 °C (-40 … +248 °F)
1 h: 140 °C/284 °F cleaning tem-
perature
-40 … +150 °C (-40 … +302 °F)
EPDM (ET 7056)
-40 … +120 °C (-40 … +248 °F)
1 h: 140 °C/284 °F cleaning tem-
perature
-
FFKM (Kalrez 6375)
-20 … +120 °C (-4 … +248 °F)
-20 … +150 °C (-4 … +302 °F)
FFKM (Perlast G75S)
-15 … +120 °C (-4 … +248 °F)
-15 … +150 °C (5 … +302 °F)
FFKM (Perlast G75B)
-15 … +120 °C (-4 … +248 °F)
-15 … +150 °C (5 … +302 °F)
Vibration resistance
mechanical vibrations with 4 g and 5 … 100 Hz
13)
Shock resistance
Acceleration 100 g/6 ms
14)
Electromechanical data - version IP 66/IP 67
Cable entry/plug
15)
12)
With process fitting PVDF, max. 100 °C (212 °F).
13)
Tested according to the guidelines of German Lloyd, GL directive 2.
14)
Tested according to EN 60068-2-27.
15)
Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.