11 Supplement

VEGAPULS 68 • Foundation Fieldbus

36538-EN-121011

Tracking speed of the measuring window

max.

1 m/min

Beam angle

2)

Ʋ Horn antenna ø 40 mm (1.575 in)

20°

Ʋ Horn antenna ø 48 mm (1.89 in)

15°

Ʋ Horn antenna ø 75 mm (2.953 in)

10°

Ʋ Horn antenna ø 95 mm (3.74 in)

Ʋ Parabolic antenna

Emitted HF power (depending on the parameter adjustment)

3)

Ʋ Average spectral transmission power

density

-14 dBm/MHz EIRP

Ʋ Max. spectral transmission power

density

+43 dBm/50 MHz EIRP

Ʋ Max. power density in a distance of

1 m

< 1 µW/cm²

Ambient conditions
Ambient, storage and transport tempera-

ture

-40 … +80 °C (-40 … +176 °F)

Process conditions
For the process conditions, please also note the specifications on the type label. The lowest value

always applies.

Seal

Antenna impedance

cone

Process temperature (measured on the process

fitting)

FKM (SHS FPM 70C3

GLT)

PTFE

-40 … +130 °C (-40 … +266 °F)

PTFE

4)

-40 … +200 °C (-40 … +392 °F)

PEEK

5)

-40 … +200 °C (-40 … +392 °F)

FFKM (Kalrez 6375)

PTFE

-20 … +130 °C (-4 … +266 °F)

PEEK

-20 … +250 °C (-4 … +482 °F)

FFKM (Kalrez 2035)

PTFE

-15 … +130 °C (5 … +266 °F)

PEEK

-15 … +210 °C (5 … +410 °F)

FFKM (Kalrez 6230)

PTFE

-15 … +130 °C (5 … +266 °F)

PEEK

-15 … +250 °C (5 … +482 °F)

Graphite

Ceramic

-200 … +450 °C (-328 … +842 °F)

Graphite (process fitting

Hastelloy C 22)

Ceramic

-196 … +400 °C (-321 … +752 °F)

Vessel pressure - horn antenna

Ʋ Antenna impedance cone PTFE

-1 … 40 bar (-100 … 4000 kPa/-14.5 … 580 psig)

Ʋ Antenna impedance cone PP

-1 … 3 bar (-100 … 300 kPa/-14.5 … 43.5 psig)

2)

Outside the specified beam angle, the energy of the radar signal has a level which is reduced by 50 % (-3 dB)

3)

EIRP: Equivalent Isotropic Radiated Power

4)

Not with steam