beautypg.com

VEGA VEGAPULS 68 (≥ 2.0.0 - ≥ 4.0.0) Profibus PA User Manual

Page 57

background image

57

9 Diagnosis, Asset Management and service

VEGAPULS 68 • Profibus PA

36537-EN-130205

Code
Text mes-

sage

Cause

Rectification

PA Dev-

Spec
Diagnosis

F125
Unper-

missible

electronics

tempera-

ture

– Temperature of the

electronics in the non-

specified section

– Check ambient tempe-

rature

– Isolate electronics

– Use instrument with

higher temperature

range

Bit 8

F260
Error in the

calibration

– Error in the calibration

carried out in the

factory

– Error in the EEPROM

– Exchanging the elec-

tronics

– Send instrument for

repair

Bit 9

F261
Error in the

configura-

tion

– Error during setup

– False signal suppres-

sion faulty

– Error when carrying

out a reset

– Repeat setup

– Repeat reset

Bit 10

F264
Installati-

on/Setup

error

– Adjustment not within

the vessel height/

measuring range

– Max. measuring range

of the instrument not

sufficient

– Check or correct

installation and/or

parameter adjustment

– Use an instrument

with bigger measuring

range

Bit 11

F265
Measu-

rement

function

disturbed

– Sensor no longer

carries out a measu-

rement

– Operating voltage

too low

– Check operating

voltage

– Carry out a reset

– Separate operating

voltage briefly

Bit 12

The following table shows the error codes and text messages in the

status message "Function check" and provides information on causes

as well as corrective measures.

Code
Text mes-

sage

Cause

Rectification

C700
Simulation ac-

tive

– A simulation is active

– Finish simulation

– Wait for the automatic end

after 60 mins.

The following table shows the error codes and text messages in the

status message "Out of specification" and provides information on

causes as well as corrective measures.

Code
Text mes-

sage

Cause

Rectification

S600
Unpermissib-

le electronics

temperature

– Temperature of the elect-

ronics in the non-specified

section

– Check ambient temperature

– Isolate electronics

– Use instrument with higher

temperature range

Function check

Out of specification