beautypg.com

VEGA VEGABAR 55 4 … 20 mA_HART - climate compensated User Manual

Page 56

background image

56

10 Supplement

VEGABAR 55 • 4 … 20 mA/HART - climate compensated

39441-EN-130321

Pressure stage, process fitting

Ʋ Thread 316L, depending on connec-

tion

PN 25, PN 60

Ʋ Thread Aluminium

PN 25

Ʋ Hygienic fittings 316L, depending on

connection

PN 6, PN 10, PN 25, PN 40 (PN 40 only with DRD and

DIN 11851)

Ʋ Flange 316L

PN 16/PN 40 or 150 lb/300 lb

Product temperature

Ʋ Standard

-12 … +140 °C (+10 … +284 °F)

Ʋ with extenation, extended thread or

Clamp 2½"

-12 … +140 °C (+10 … +284 °F)

Ʋ with cooling element

-12 … +180 °C (+10 … +356 °F)

Ʋ with cooling element and screening

sheet

-12 … +200 °C (+10 … +392 °F)

Vibration resistance

mechanical vibrations with 4 g and 5 … 100 Hz

12)

Shock resistance

Acceleration 100 g/6 ms

13)

Electromechanical data - version IP 66/IP 67
Cable entry/plug

14)

Ʋ Single chamber housing

– 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT

or:

– 1␣x plug (depending on the version), 1␣x blind stopper

M20␣x␣1.5

or:

– 2 x blind plug M20 x 1,5

Ʋ Double chamber housing

– 1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5; plug M12 x 1 for the external

display and adjustment unit (optional)

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT, plug

M12 x 1 for the external display and adjustment unit

(optional)

or:

– 1 x plug (depending on the version), 1 x blind plug

M20 x 1.5; plug M12 x 1 for the external display and

adjustment unit (optional)

or:

– 2 x blind stoppers M20 x 1.5; plug M12 x 1 for the

external display and adjustment unit (optional)

12)

Tested according to the guidelines of German Lloyd, GL directive 2.

13)

Tested according to EN 60068-2-27.

14)

Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.