VEGA VEGABAR 55 4 … 20 mA_HART - climate compensated User Manual
Page 56
56
10 Supplement
VEGABAR 55 • 4 … 20 mA/HART - climate compensated
39441-EN-130321
Pressure stage, process fitting
Ʋ Thread 316L, depending on connec-
tion
PN 25, PN 60
Ʋ Thread Aluminium
PN 25
Ʋ Hygienic fittings 316L, depending on
connection
PN 6, PN 10, PN 25, PN 40 (PN 40 only with DRD and
DIN 11851)
Ʋ Flange 316L
PN 16/PN 40 or 150 lb/300 lb
Product temperature
Ʋ Standard
-12 … +140 °C (+10 … +284 °F)
Ʋ with extenation, extended thread or
Clamp 2½"
-12 … +140 °C (+10 … +284 °F)
Ʋ with cooling element
-12 … +180 °C (+10 … +356 °F)
Ʋ with cooling element and screening
sheet
-12 … +200 °C (+10 … +392 °F)
Vibration resistance
mechanical vibrations with 4 g and 5 … 100 Hz
12)
Shock resistance
Acceleration 100 g/6 ms
13)
Electromechanical data - version IP 66/IP 67
Cable entry/plug
14)
Ʋ Single chamber housing
– 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x
blind plug M20 x 1.5
or:
– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT
or:
– 1␣x plug (depending on the version), 1␣x blind stopper
M20␣x␣1.5
or:
– 2 x blind plug M20 x 1,5
Ʋ Double chamber housing
– 1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm), 1 x
blind plug M20 x 1.5; plug M12 x 1 for the external
display and adjustment unit (optional)
or:
– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT, plug
M12 x 1 for the external display and adjustment unit
(optional)
or:
– 1 x plug (depending on the version), 1 x blind plug
M20 x 1.5; plug M12 x 1 for the external display and
adjustment unit (optional)
or:
– 2 x blind stoppers M20 x 1.5; plug M12 x 1 for the
external display and adjustment unit (optional)
12)
Tested according to the guidelines of German Lloyd, GL directive 2.
13)
Tested according to EN 60068-2-27.
14)
Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.