VEGA VEGAMIP R61 Receiving unit - Transistor User Manual
Page 38
-
VEGAMIP
61, horn antenna/316L
-1 … 40 bar/-100 … 4000 kPa (-14.5 … 580 psig)
-
VEGAMIP
61, encapsulated horn anten-
na with PTFE cover
-1 … 16 bar/-100 … 1600 kPa (-14.5 … 232 psig)
-
VEGAMIP
61 with mounting adapter
unpressurized (IP 67)
Process temperature (thread or flange tem-
perature)
-
VEGAMIP
61, threaded version - internal
horn antenna with PTFE cover
-40 … +80 °C (-40 … +176 °F)
-
VEGAMIP
61, plastic encapsulated an-
tenna with PP cover
-40 … +80 °C (-40 … +176 °F)
-
VEGAMIP
61, horn antenna/316L - seal:
FKM (SHS FDM 70C3 GLT)
-40 … +130 °C (-40 … +266 °F)
-
VEGAMIP
61, horn antenna/316L - seal:
FFKM (Kalrez 6375)
-20 … +130 °C (-4 … +266 °F)
-
VEGAMIP
61, encapsulated horn anten-
na with PTFE cover
-40 … +200 °C (-40 … +392 °F)
-
VEGAMIP
61 with mounting adapter
150 mm (optional)
-40 … +250 °C (-40 … +482 °F)
-
VEGAMIP
61 with mounting adapter
300 mm (optional)
-40 … +450 °C (-40 … +842 °F)
Electromechanical data
Cable entry/plug (dependent on the version)
-
Single chamber housing
l
1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5; attached 1 x cable
entry M20 x 1.5
or:
l
1 x cable entry ½ NPT, 1 x blind stopper ½ NPT,
1 x cable entry ½ NPT
or:
l
1 x plug M12 x 1; 1 x blind stopper M20 x 1.5
Spring-loaded terminals
for wire cross-section up to 1.5 mm² (AWG 16)
Voltage supply
Operating voltage
20 … 55 V DC
Power consumption
max. 0.8 W
Electrical protective measures
Protection rating
IP
66/IP 67
Overvoltage category
III
Protection class
II
Approvals
Instruments with approvals can have different technical data depending on the version.
38
VEGAMIP R
61 • - Transistor
9 Supplement
40694
-EN