beautypg.com

7 maintenance and fault rectification, 1 maintaining, 2 remove interferences – VEGA VEGAMIP R61 Receiving unit - Transistor User Manual

Page 33: 7maintenance and fault rectification

background image

7

Maintenance and fault rectification

7

.1 Maintaining

If the device is used correctly, no maintenance is required in normal
operation.

7

.2 Remove interferences

The operator of the system is responsible for taking suitable measures
to rectify faults.

The first measure to be taken is to check the output signal. In many
cases, the causes can be determined this way and the faults rectified.

Error

Cause

Rectification

VEGAMIP

61

signals

"

covered" without being

submerged (overfill pro-
tection)
or

VEGAMIP

61

signals

"

uncovered" when being

submerged (dry run pro-
tection)

Operating voltage too low

Check operating voltage

Buildup on the sensor

Check if there is buildup on the two sensors, and if so,
remove it.

Wrong mode selected

Set the correct mode on the mode switch (max.:
overflow protection, min.: dry run protection). Wiring
should be carried out according to the quiescent current
principle.

Unfavourable installation
location

Mount the sensor (emitter/receiver) in a position where
there is a lot of measured product between emitting and
receiving unit.

Mount the sensor (emitter/receiver) in a psoition where
there are no interfering installations or moving parts.

Electronics defective

Push the mode switch. If the instrument then does not
change the mode, the electronics module may be
defective. Exchange the electronics module.

Sensor defective

Push the mode switch (min./max.). If the instrument
switches over, the sensor can be covered or mechani-
cally damaged.
Should the switching function on the correct mode be
wrong again, the you should return the instrument for
repair.

Signal lamp lights red

Operating voltage too low

Check operating voltage

Electronics module has
detected an internal failure

Exchange the instrument or send it in for repair

Instrument switches de-
layed

Check switching delay

Adjust switching delay correctly

Reaction when malfunc-
tions occur

Fault rectification

Checking the switching
signal

VEGAMIP R

61 • - Transistor

33

7 Maintenance and fault rectification

40694

-

EN

-120228