Information:
Make sure that the housing is not rotated during the electronics
exchange. Otherwise the plug may be in a different position later.

8

Screw in the two holding screws and tighten them

9

Insert the terminal block again. You must hear it snap in.

10 Screw the housing cover back on

The electronics exchange is now finished.

As a rule, an exchange of electronics must be documented internally
when Ex applications are involved.

To exchange the electronics module, proceed as follows:

1

Switch off power supply

2

Unscrew the housing cover

3

Lift the terminal blocks with a small screwdriver and pull them out.

4

Loosen the two screws with a screw driver (Torx size T10 or slot 4)

2

1

Fig. 2: Loosening the holding screws

1

Electronics module

2

Screws (2 pcs.)

5

Pull out the old electronics module

6

Compare the new electronics module with the old one. The type
label of the electronics module must correspond to that of the old
electronics module. This applies particularly to instruments used in
hazardous areas.

7

Insert the electronics module carefully. Make sure that the plug is
in the correct position.

Information:
Make sure that the housing is not rotated during the electronics
exchange. Otherwise the plug may be in a different position later.

8

Screw in the two holding screws and tighten them

9

Insert the terminal blocks again. You must hear them snap in.

Electronics module - re-
ceiving unit MP60EE.R**

8

Electronics module • VEGAMIP series 60 - receiving unit

4 Mounting

37349

-EN

-121205