VEGA VEGAWELL 72 4 … 20 mA_HART PT 100 User Manual
Page 36

Total deviation (similar to DIN 16086)
T
he total deviation F
t
,
also called practical deviation, is the sum of the basic accuracy F
p
and long-term stability:
F
t
= F
p
+ F
s
F
perf
= √((F
T
)
2
+ (F
K
l
)
2
)
W
ith
-
F
t
: F
total
,
total deviation
-
F
p
: F
perf
,
basic accuracy
-
F
s
: F
stab
,
long-term drift
-
F
T
: T
emperature coefficient (influ-
ence of medium or ambient tem-
perature)
-
F
K
l
: D
eviation
Ambient conditions
A
mbient temperature
-
C
onnection cable PE
-40 … +60 °C (-40 … +140 °F)
-
C
onnection cable PUR, FEP
-40 … +80 °C (-40 … +176 °F)
S
torage and transport temperature
-40 … +80 °C (-40 … +176 °F)
Process conditions
M
ax. process pressure, transmitter
8)
-
M
easuring range 0.1 bar (1.45 psig)
15
bar (218 psig)
-
M
easuring range 0.2 bar (2.9 psig)
20
bar (290 psig)
-
M
easuring range ≤ 0.4 bar (5.8 psig)
25
bar (363 psig)
P
ressure stage, process fitting
-
U
nassembled threaded fitting
316L: PN 3, PVDF:
unpressurized
-
T
hread on the housing
PN
3
P
roduct temperature, depending on the version
Suspension cable
Measuring cell seal
Product temperature
PE
FKM
-20 … +60 °C (-4 … +140 °F)
PUR
FKM
-20 … +80 °C (-4 … +176 °F)
PUR
FFKM
-20 … +80 °C (-4 … +176 °F)
FEP
FFKM
-10 … +80 °C (+14 … +176 °F)
8)
L
imited by the overpressure resistance of the measuring cell.
36
VEGAWELL
72 • 4 … 20
mA/HART PT
9 S
upplement
34360
-EN
-080806